Quick Results Guide 9
 
NOTE
 
Ohms guard cannot be used for high current (>100mA) ranges. Ohms guard cannot 
be selected if already on a high current range. Conversely, if ohms guard is already 
selected, a high current range cannot be selected.
 
The following menu sequence selects ohms guard:
Press CONFIG > press MEAS 
 
Ω
 
 (or SOURCE V or SOURCE I) > select GUARD > select 
OHMS
Basic connections for ohms guard is shown in Figure 6A. With V, 
 
Ω
 
 Guard and In/Out HI at 
virtually the same potential, the voltage drop across R will be 0V, and therefore, the current 
through R will be 0A. Virtually all the current from In/Out HI will flow through the DUT 
resulting in an accurate measurement.
If the guard resistance path (R
 
G
 
) is <1k
 
Ω
 
, IR drop in the V, 
 
Ω
 
 Guard test lead could be high 
enough such that a significant voltage drop appears across R. The resultant leakage current 
through R will corrupt the measurement of the DUT.
Guard Sense is used to cancel the effect of IR drop in the V, 
 
Ω
 
 Guard test lead as shown in 
Figure 6B. Guard Sense regulates the guard voltage to ensure that the same potential appears 
on either side of R.
For DUT <1k
 
Ω
 
, 4-wire sensing should be used as shown in Figure 6C.
 
NOTE
 
Guard current (I
 
G
 
) must never exceed 50mA. If it does, guard voltage will become 
less than the output voltage and corrupt the measurement.
DUT
V, Ω Guard
Sense HI
In/Out HI
In/Out LO
Sense LO
Guard Sense
Sense Mode: 4-wire
C.  6-wire connections
DUT
V, Ω Guard
In/Out HI
In/Out LO
Sense Mode: 2-wire
A.  Basic connections
DUT
V, Ω Guard
In/Out HI
In/Out LO
Guard Sense
Sense Mode: 2-wire
B.  Connections using guard sense
<1kΩ
RG
IG
≥1kΩ
RG
IG
R
0V
R
0V
Figure 6
Guarded ohms measurements