12 Quick Results Guide
 
Measure only (V or I)
 
The SourceMeter can be used to measure voltage only (voltmeter) or current only (amme-
ter). The procedure to measure only is provided in Table 3.
 
CAUTION 1
 
When using the SourceMeter as a voltmeter only, voltage compliance must 
be set higher than the voltage that is being measured. Failure to do so 
could result in instrument damage due to excessive current that will flow 
into the SourceMeter.
CAUTION 2 When using the SourceMeter as a voltmeter only, DO NOT use auto range 
and NEVER select a measurement range that is below the applied signal 
level. Otherwise, high current drawn from the external source could dam-
age the external source or the test circuit.
Table 3
Measure only (V or I) procedure
Procedure Details
1.  Select source-measure 
functions.
To measure voltage, press SOURCE I and MEAS V. To mea-
sure current, press SOURCE V and MEAS I.
2.  Set source level to zero. Use DISPLAY EDIT key to place cursor in source field (Vsrc 
or Isrc), use RANGE 
 key to select lowest range, press 
MENU to display zero source level, then press ENTER.
3.  Set compliance higher than 
the expected measurement 
(see CAUTION 1).
Use DISPLAY EDIT key to place cursor in Cmpl field, use 
RANGE 
 and  keys to select range, use edit keys to key 
in compliance value, then press ENTER.
4.  Select measurement range 
(see Note).
Use RANGE  and  keys to select a fixed measurement 
range. For best accuracy, use the lowest possible range.
5.  Connect voltage or current 
to be measured.
Connect DUT to SourceMeter using 2-wire connections (see 
Figure 3).
6.  Turn output on and take 
reading from display.
Press ON/OFF OUTPUT key. Red OUTPUT indicator on.
7.  Turn output off when 
finished.
Press ON/OFF OUTPUT key. Red OUTPUT indicator off.
Note: When measuring current, auto range can be used. When measuring voltage, DO NOT use auto range 
(see CAUTION 2).