22 Quick Results Guide
Remote command programming
The SCPI commands for speed, digits, filter, and rel are listed in Table 12. The commands for 
ranging are listed in Tables 6 and 7.
Table 12
SCPI commands; speed, digits, filter, and rel
Command Description
 Speed commands
 1
:   
  [:SENSe]:CURRent:NPLCycles  <n> Set measurement speed; <n> = 0.01 to 10. 
  [:SENSe]:VOLTage:NPLCycles  <n> Set measurement speed; <n> = 0.01 to 10. 
  [:SENSe]:RESistance:NPLCycles  <n> Set measurement speed; <n> = 0.01 to 10. 
Digits command:
:DISPlay:DIGits  <n> Set display resolution; <n> = 4, 5, 6, or 7. 
Filter commands: 
[:SENSe]:AVERage:TCONtrol <name> Select filter type; <name> = REPeat or MOVing.
[:SENSe]:AVERage:COUNt  <n> Set filter count; <n> = 1 to 100. 
[:SENSe]:AVERage  <b> Enable/disable filter; <b> = ON or OFF. 
Rel command 
2
:
:CALCulate2:NULL:STATe  <b> Enable/disable rel; <b> = ON or OFF. 
1. The speed setting is global for all functions. Therefore, you can use any of the three commands to set speed. 
2. The commands to acquire a rel value are not listed in this table (see Section 8, Relative in the 2400 Series 
SourceMeter User’s Manual).