Table 16-2
Model 2601/2602 calibration steps
Function Calibration steps
2
Calibration points
4
Sense mode
5
Voltage Source
and Measure
1
100mV
100mV
1V
1V
6V
40V
±1e-10, ±90mV
±1e-10, ±90mV
±1e-10, ±0.9V
±1e-10, ±0.9V
±1e-10, ±5.4V
±1e-10, ±36V
smuX.SENSE_LOCAL
smuX.SENSE_REMOTE
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
Current Source
and Measure
1
100nA
1μA
10μA
100μA
1mA
1mA
10mA
100mA
1A
3A
10A
3
±1e-10, ±90nA
±1e-10, ±0.9μA
±1e-10, ±9μA
±1e-10, ±90μA
±1e-10, ±0.9mA
±1e-10, ±0.9mA
±1e-10, ±9mA
±1e-10, ±90mA
±1e-10, ±0.9A
±1e-10, ±2.4A
±1e-10, ±2.4A
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
1. Calibrate only the source for the SENSE_CALA sense steps.
2. Steps must be performed in the order shown.
3. 10A range for changing calibration of range only and is not available for normal use.
4. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-10. Calibration
polarities must also be set as shown in the procedures.
5. Output must be off before changing to the CALA sense mode.
16-6 Calibration Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006 Return to Section 16 topics