Table 16-3
Model 2611/2612 calibration steps
Function Calibration steps
2
Calibration points
3
Sense mode
4
Voltage Source
and Measure
1
200mV
200mV
2V
2V
20V
200V
±1e-10, ±180mV
±1e-10, ±180mV
±1e-10, ±1.8V
±1e-10, ±1.8V
±1e-10, ±18V
±1e-10, ±180V
smuX.SENSE_LOCAL
smuX.SENSE_REMOTE
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
Current Source
and Measure
1
100nA
1μA
10μA
100μA
1mA
1mA
10mA
100mA
1A
1.5A
10A
±1e-10, ±90nA
±1e-10, ±0.9μA
±1e-10, ±9μA
±1e-10, ±90μA
±1e-10, ±0.9mA
±1e-10, ±0.9mA
±1e-10, ±9mA
±1e-10, ±90mA
±1e-10, ±0.9A
±1e-10, ±1.35A
±1e-10, ±2.4A
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_CALA
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
smuX.SENSE_LOCAL
1. Calibrate only the source for the SENSE_CALA sense steps.
2. Steps must be performed in the order shown.
3. Do not use actual 0 values for zero calibration points. Send very small values such as ±1e-10. Calibration
polarities must also be set as shown in the procedures.
4. Output must be off before changing to the CALA sense mode.
Series 2600 System SourceMeters Reference Manual Calibration 16-7
Return to Section 16 topics 2600S-901-01 Rev. A / May 2006
Calibration commands
Table 16-4 summarizes remote calibration commands. For a more complete
description of these commands, refer to Section 12.