B-8 Frequently Asked Questions Series 2600 System SourceMeters User’s Manual
2600S-900-01 Rev. A / May 2006 Return to In this Appendix:
The three measure readings, timestamps, and source values will be stored in the
buffer.
Step 3: Read the buffer
The following code will output the three sets of data (nine values) in the same
message:
rb1 = smua.nvbuffer1
printbuffer(1, rb1.n, rb1, rb1.timestamps, rb1.sourcevalues)
For the above printbuffer function, 1 is the starting index for values to output;
rb1.n is the ending index (n = the number of readings in the buffer, which in this
case is 3);
rb1 includes the measured readings; rb1.timestamps includes the
timestamps; and
rb1.sourcevalues includes the source values.
Example output:
1.234567e-03, 5.555555e-02, 1.000000e+00, 2.362360e-03,
8.888888e-02, 1.000000e+00, 2.362368e-03, 1.111111e-01,
1.000000e+00
The above output is returned in the following order:
reading1, timestamp1, sourcevalue1, reading2, timestamp2, sourcevalue2,
reading3, timestamp3, sourcevalue3
Refer to Step 3 of “Running a factory script” in Section 3 for more examples on
reading the buffer.

Table of Contents

Other manuals for Keithley SourceMeter 2602

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2602 and is the answer not in the manual?

Keithley SourceMeter 2602 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2602
CategoryMeasuring Instruments
LanguageEnglish

Summary

Warranty Details

Product Warranty Terms

Details Keithley's product warranty terms and conditions for defects.

Warranty Exclusions and Limitations

Defines exclusions and limitations to the product warranty, including misuse and consequential damages.

Instrument Safety Precautions

Electrical Safety and Hazard Warnings

Covers shock hazards, voltage precautions, and safe practices for electrical measurements.

Front Panel Operations Guide

Front Panel Controls and Setup

Details front panel keys, switches, and settings for source-measure operations.

Source-Measure Specifications

Lists the DC source-measure specifications for Models 2601/2602 and 2611/2612.

Performing Basic Measurements

Procedure for setting up the instrument and performing basic measurements.

Remote Operation Guide

Remote Interface Configuration

Connects the instrument to a computer via GPIB or RS-232 for remote control.

TSB Measurements Procedure

Demonstrates using the TSB Instrument Console to send commands for measurements.

TSP-Link System Control

TSP-Link System Connection and Setup

Connects multiple instruments to form a system and assign node numbers.

Related product manuals