Series 2600 System SourceMeters Reference Manual System Expansion (TSP-Link) 9-7
Return to Section 9 topics 2600S-901-01 Rev. A / May 2006
System behavior
Using the reset() command
While most TSP-Link operations target a single node in the system, the reset()
command affects the system as a whole. The
reset() command, by definition,
resets all nodes to their default settings:
reset() -- Resets all nodes in a TSP-Link system.
node[N] and localnode can be used with reset to reset only one of the nodes.
The other nodes are not affected. Examples:
node[1].reset() -- Resets Node 1 only.
localnode.reset() -- Resets Node 1 only.
node[4].reset() -- Resets Node 4 only.
Abort
An abort will terminate an executing script and return all nodes to local operation
(REM annunciators turn off), dissolving the Master/Slave relationships between
nodes. An abort is invoked by either issuing an
abort command to the Master or
pressing the EXIT key on any node in the system.
An abort can also be performed by pressing the OUTPUT ON/OFF key on any
node. The results are the same as above, with the addition that all SMU outputs in
the system are turned off.

Table of Contents

Other manuals for Keithley SourceMeter 2611

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley SourceMeter 2611 and is the answer not in the manual?

Keithley SourceMeter 2611 Specifications

General IconGeneral
BrandKeithley
ModelSourceMeter 2611
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals