Carry out adjustment as near as possible to the balance’s maximum weight
(recommended adjustment weight see chap. 1). Weights of different nominal
values or tolerance classes may be used for adjustment but are not optimal
for technical measuring. The accuracy of the adjustment weight must corres-
pond approximately to or, if possible, be better than, the readability [d] of the
balance. Info about test weights can be found on the Internet at:
http://www.kern-sohn.com
Observe stable environmental conditions. A warm up time (see chapter 1) is
required for stabilization.
Ensure that there are no objects on the weighing plate.
Avoid vibration and air flow.
Always carry out adjustment with the standard weighing plate in place.