6
.3 Backup
(
B
K
UP)
O
pe
r
a
tions
(1) In the service menu, select [NVRAM]
[BKUP (Backup)].
(2) The execution confirmation message "OK?" appears.
(3) Press the [Enter] key. Data in the EEPROM is backed up and saved to the microprocessor flash
memory.
(4) The backup progress indicator appears.
(5) When backup is finished, the screen returns to [NVRAM] in the service menu.
6
.4 D
o
wnlo
a
d
(
RES
T
R
)
O
pe
r
a
tions
(1) In the service menu, select [NVRAM]
[RESTR] (download).
(2) The execution confirmation message "OK?" appears.
(3) Press the [Enter] key. Data backed up in the microprocessor flash memory is transferred to the
EEPROM.
(4) The download progress indicator appears.
(5) When downloading is finished, the screen returns to [NVRAM] in the service menu.
The following items are not downloaded in RESTR. Setup is required.
Serial Number
Factory setting(SHIP)
Setting information Lock(LOCK)
OEM model setting(OEM Model Only.)
6
.5 All Initi
a
liz
a
ti
o
n
(
INI
T
)
O
p
e
r
a
tions
(1) In the service menu, select [NVRAM]
[INIT] (initialize all).
(2) The execution confirmation message "OK?" appears.
(3) Press [Enter] to confirm execution.
(4) The initialization progress indicator appears.
(5) When initialization is finished, the screen returns to [NVRAM] in the service menu.
52 DBS-SH-e-1210

Table of Contents

Other manuals for KERN DBS 60-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN DBS 60-3 and is the answer not in the manual?

KERN DBS 60-3 Specifications

General IconGeneral
BrandKERN
ModelDBS 60-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals