7.
5
T
empe
r
a
tu
r
e
Calibrate the temperature using the same operations as in the user mode and confirm
the value indicated on the temperature calibration kit.
T
a
ble 1 DBS Inspecti
o
n
S
tand
a
r
ds
Calibration
Error
180
°
C
Model
DBS 60-3
Weighing
Capacity 60
g
Minimum
Val ue
Displayed 0.001
g
Repeatability
(Rx,
Ry)
Calibration
Weight
0.006
g
50
g
Cornerload
Weight
±
0.002
g
20
g
Linearity
±
0.005
g
Tilt
Error
±
0.03
g
Temperature
100
°
C
±
5
°
C