9.
6
Does N
o
t
R
each
T
empe
r
a
tu
r
e Setting
Does not
r
each tem
p
e
r
a
tu
r
e setting.
Y
es
Dirty
or
contaminated?
No
No
Was
temperature
calibrated
by
the
user?
Y
es
No
Is the cumulative
time less than 5000
hours?
Y
es
Is the resistance
greater
than 100kΩ ?
Y
es
No
Replace
the thermistor.
Replace
the heater.
Check
the
heater
(S5).
Replace
the heater.
Open
the
service menu and
check
the heater time
count.
Calibrate temperature
(user
mode).
Clean
the
heater
or
glass
holder.
Check for
a
dirty heater
or
glass
holder.