measurement.
"ERR.121" is displayed.
"ERR.122" is displayed.
"ERR.123" is displayed.
3) Dirty heater (S5)
Remove the dirt/contamination.
However, do not touch directly with
hands.
4) Dirty glass holder Remove the contamination.
5) Temperature calibration is incorrect. Recalibrate temperature.
"ERR.200" is displayed.
Power supply voltage does not match the
factory setting.
Supply the correct voltage.
Modify the factory setting.
"ERR.202" is displayed.
"ERR.201" is displayed.
1) Unstable power supply frequency or
voltage.
Supply the correct frequency and
voltage.
2) Defective power board assembly (B4)
Replace the power board assembly
(B4).
3) Defective main board assembly (B2)
Replace the main board assembly
(B2).
"COM.ERR" is displayed.
1) Defective power board assembly (B4)
Replace the power board assembly
(B4).
2) Defective main board assembly (B2)
Replace the main board assembly
(B2).
"ERR.124" is displayed.
See "Zero-drift or span-drift is large" above.
See "Reproducibility for weight measurements is poor" above.
See "Does not reach temperature specified during moisture level measurement"
above.
DBS-SH-e-1210 57

Table of Contents

Other manuals for KERN DBS 60-3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN DBS 60-3 and is the answer not in the manual?

KERN DBS 60-3 Specifications

General IconGeneral
BrandKERN
ModelDBS 60-3
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals