EMS-BA-e-1716 11
5.6 Linearization
(only models EMS 300-3, EMS 3000-2)
Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a monitoring of test resources, you can
improve this by means of linearization.
well acquainted with the workings of weighing scales.
The test weights to be used must be adapted to the weighing scale’s
specifications; see chapter 3.4 “Testing instruments control”.
Observe stable environmental conditions. Stabilisation requires a
certain warm-up time.
After successful linearization you will have to carry out calibration;
Tab. 1: Adjustment points
Adjustment
weight
EMS 300-3 EMS 3000-2
1.
50 g
500 g
2.
100 g
1000 g
3.
150 g
1500 g
4.
200 g
2000 g
5.
300 g
3000 g
Operation
Display
How to carry out linearization:
Switch on balance
Press repeatedly until AF is
displayed

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN EMS Series and is the answer not in the manual?

KERN EMS Series Specifications

General IconGeneral
Weighing plate materialstainless steel
CalibrationExternal calibration
DisplayLCD
Overall dimensionsVaries by model
Permissible ambient temperature10 °C - 40 °C
InterfaceRS-232
Protection ClassIP54 (optional)

Related product manuals