8 EOC-BA-e-1920
KERN EOC 60K-3A EOC 60K-3L EOC 100K-2
Readability (d) 5 g 1 g / 2 g 20 g / 50 g
Weighing range (max) 60 kg 30 kg / 60 kg 60 kg / 150 kg
Reproducibility 5 g 1 g / 2 g 20 g / 50 g
Linearity ± 15 g ± 3 g / 6 g ± 60 / 150 g
Warm-up time 30 minutes 2 hours 10 minutes
Minimum unit weight at
piece counting under
laboratory conditions*
2 g 2 g 5 g
Minimum unit weight at
piece counting under
normal conditions**
20 g 20 g 50 g
Reference unit weights at
piece count
10, 20, 50, 100, 200
Weighing Units Details Weighing units“ chapter 7.6
Recommended adjustment
weight, not added (class)
Details for „Selection of
the Adjustment weight“
in chapter 7.6
60 kg
(F2)
60 kg
(M1)
150 kg
(M1)
Stabilization time (typical) 2 sec.
Electric Supply
100 V - 240 V, 50 / 60 Hz
Auto Off
off, 3 min., 5 min., 15 min., 30 min.
Operating temperature - 10° C .... + 40° C
Humidity of air max. 80 % (not condensing)
Terminal (B x D x H) mm 268 x 115 x 80
Platform (B x D x H )mm
300 x 300 x 110 500 x 400 x 120 300 x 300 x 110
Weight kg (net) 5.2 9.0 5.2