Servicemanual for electronic precision
balances series
KERN EW/EG
Version 1.1 07/98
EW/EG-SU-e-9811

Questions and Answers

  • T
    Tracy WilliamsAug 16, 2025
    Why do the KERN EW - 300 - 2M Scales display not light up?
    • O
      owenssherylAug 16, 2025
      The display might not light up due to several reasons: * A defective EW/EG board – consider replacing it. * A defective AC adapter – consider replacing it. * Incorrectly connected cords inside – check and correct their connections. * A discharged built-in battery (if applicable) – consider replacing it.
  • F
    Fred TrujilloAug 19, 2025
    What to do if the KERN EW - 300 - 2M Scales display doesn't settle down or the zero point drifts?
    • S
      Stephen ButlerAug 19, 2025
      If the display doesn't settle, repeat correctly, or the zero point drifts, it could be due to: * Some parts, like the stopper, touching others – adjust the position of the stopper. * The weighing pan touching other parts – ensure it doesn't touch anything. * A foreign substance inside – remove it. * A defective tuning-fork sensor or mechanism unit – consider replacing it. * Environmental factors like wind or oscillation – check the environment, working base, and stabilization time setting.
  • D
    destiny62Aug 22, 2025
    What does o-Err mean on KERN Scales when the load is less than specified?
    • J
      jordancallahanAug 22, 2025
      An 'o-Err' message with a net load less than specified capacity could mean: * The weight on the scale pan (net weight + tare value) exceeds the scale capacity – reduce the weight. * Coefficient memories have changed due to noise or static electricity – adjust linearity. * There's a mistake in the setting of address data – correct the setting. * The calibration weight is incorrect – use the correct calibration weight.
  • R
    Robert GutierrezAug 25, 2025
    What to do if the display on my KERN EW - 300 - 2M suddenly disappeared?
    • D
      Donald HermanAug 25, 2025
      If the display suddenly disappears during operation, it could be due to: * A consumed built-in battery (if applicable) – consider replacing it. * A defective EW/EG board – consider replacing it. * A defective AC adapter – consider replacing it.
  • M
    matthewrobinsonAug 27, 2025
    What does 'u-Err' or 'o-Err' after self-test mean on KERN EW - 300 - 2M?
    • K
      Kristine MurphyAug 27, 2025
      If you see 'u-Err' or 'o-Err' after the self-test, it might be caused by: * Using the wrong weighing pan – use the correct one. * A defective tuning-fork sensor or mechanism unit – consider replacing it. * A defective EW/EG board – consider replacing it. * A mistake in the setting of address data – correct the setting. * Coefficient memories changed by noises or static electricity – adjust linearity.
  • Z
    zgeorgeAug 30, 2025
    Why is the span out of the specified range on my KERN Scales?
    • L
      levi22Aug 30, 2025
      If the span is out of the specified range, it could be due to: * A defective tuning-fork sensor or mechanism unit – consider replacing it. * A defective EW/EG board – consider replacing it. * A mistake in the setting of address data – correct the setting. * Coefficient memories changed by noises or static electricity – adjust linearity.
  • M
    Michael Mcconnell MDSep 2, 2025
    Why is linearity out of the specified range on my KERN Scales?
    • S
      Stephanie HoldenSep 2, 2025
      If linearity is out of the specified range, it could be caused by: * A defective tuning-fork sensor or mechanism unit – consider replacing it. * A mistake in the setting of address data – correct the setting. * Coefficient memories changed by noises or static electricity – adjust linearity. * Using an incorrect calibration weight – use the correct weight.
  • R
    reneesampsonSep 5, 2025
    What to do if KERN EW - 300 - 2M Scales show too much corner error?
    • J
      John WashingtonSep 5, 2025
      If there's too much corner error, it might be because: * The mechanism unit is defective, such as a bent or twisted Roberval plate (spring) – consider replacing the mechanism unit. * The pan base is touching other parts – adjust the position of the pan base.
  • P
    patricia63Sep 7, 2025
    What does 'b-Err' mean on KERN EW - 300 - 2M?
    • F
      Frank FarrellSep 7, 2025
      A 'b-Err' message indicates that: * The EW/EG board is defective – consider replacing it. * Coefficient memories (address data) have changed due to noises or static electricity – adjust linearity.
  • K
    Kevin WoodwardSep 10, 2025
    What does '1-Err' mean on KERN EW - 300 - 2M Scales?
    • A
      Adam FordSep 12, 2025
      A '1-Err' message indicates that the calibration weight is less than ½ F.S. Use a calibration weight that is at least ½ F.S.

Summary

1 Total View of EW;WG

2 Electronic Construction

2.1 KERN EW Type Block Diagram

Block diagram for KERN EW type electronic balances.

2.2 KERN EG Type Block Diagram

Block diagram for KERN EG type electronic balances.

2.3 KERN EW Type Wiring

Wiring diagram for KERN EW type electronic balances.

2.4 KERN EG Type Wiring

Wiring diagram for KERN EG type electronic balances.

3 Troubleshooting Procedure

3.1 Troubleshooting Flowchart

Flowchart for troubleshooting electronic precision balances.

3.2 Troubleshooting Table

Table of common symptoms, causes, and remedies for balance issues.

3.3 Primary Checks

Initial checks to perform before troubleshooting balance problems.

3.4 Case Removal

Step-by-step guide on disassembling the balance case.

3.5 Lock Switch Operation

Explanation and illustration of the lock switch function.

3.6 Case Reassembly

Instructions for reassembling the balance case.

4 Adjustments and Settings

4.1 EW Type Span Calibration

Procedure for calibrating the span on EW type balances.

4.2 Corner Error Adjustment

Steps to adjust corner errors on the balance.

4.3 Overload Stopper Adjustment

Guide for adjusting the overload stopper mechanism.

4.4 EG Type Built-In Weight Calibration

Procedure for calibrating built-in weight on EG scales.

5 Parts Replacement

5.1 Mechanism Unit Replacement Sequence

Steps for replacing the weighing mechanism unit.

5.2 New PCB Installation

Procedure for installing a new DJD circuit board assembly.

6 EW Type Parts List