Servicemanual for electronic precision
balances series
KERN EW/EG
Version 1.1 07/98
EW/EG-SU-e-9811

Questions and Answers

  • J
    Julie StoneAug 17, 2025
    Why do the KERN EW - 600 - 2M display not light up?
    • M
      Michelle BennettAug 17, 2025
      The display might not light up due to several reasons: * A defective EW/EG board – consider replacing it. * A defective AC adapter – consider replacing it. * Incorrectly connected cords inside – check and correct their connections. * A discharged built-in battery (if applicable) – consider replacing it.
  • K
    kristaholtAug 19, 2025
    What to do if the KERN EW - 600 - 2M Scales display doesn't settle down or the zero point drifts?
    • W
      William MartinAug 19, 2025
      If the display doesn't settle, repeat correctly, or the zero point drifts, it could be due to: * Some parts, like the stopper, touching others – adjust the position of the stopper. * The weighing pan touching other parts – ensure it doesn't touch anything. * A foreign substance inside – remove it. * A defective tuning-fork sensor or mechanism unit – consider replacing it. * Environmental factors like wind or oscillation – check the environment, working base, and stabilization time setting.
  • N
    Nicole HillAug 22, 2025
    What does o-Err mean on KERN Scales when the load is less than specified?
    • F
      Frank WrightAug 22, 2025
      An 'o-Err' message with a net load less than specified capacity could mean: * The weight on the scale pan (net weight + tare value) exceeds the scale capacity – reduce the weight. * Coefficient memories have changed due to noise or static electricity – adjust linearity. * There's a mistake in the setting of address data – correct the setting. * The calibration weight is incorrect – use the correct calibration weight.
  • K
    Krista BassAug 25, 2025
    What to do if the display on my KERN EW - 600 - 2M suddenly disappeared?
    • B
      Benjamin GriffinAug 26, 2025
      If the display suddenly disappears during operation, it could be due to: * A consumed built-in battery (if applicable) – consider replacing it. * A defective EW/EG board – consider replacing it. * A defective AC adapter – consider replacing it.
  • T
    Tanya AustinAug 28, 2025
    What does 'u-Err' or 'o-Err' after self-test mean on KERN EW - 600 - 2M?
    • D
      Darren PierceAug 28, 2025
      If you see 'u-Err' or 'o-Err' after the self-test, it might be caused by: * Using the wrong weighing pan – use the correct one. * A defective tuning-fork sensor or mechanism unit – consider replacing it. * A defective EW/EG board – consider replacing it. * A mistake in the setting of address data – correct the setting. * Coefficient memories changed by noises or static electricity – adjust linearity.
  • B
    Brian JonesAug 30, 2025
    Why is the span out of the specified range on my KERN Scales?
    • S
      Steven ShawAug 30, 2025
      If the span is out of the specified range, it could be due to: * A defective tuning-fork sensor or mechanism unit – consider replacing it. * A defective EW/EG board – consider replacing it. * A mistake in the setting of address data – correct the setting. * Coefficient memories changed by noises or static electricity – adjust linearity.
  • Y
    yolandahansenSep 2, 2025
    Why is linearity out of the specified range on my KERN EW - 600 - 2M Scales?
    • T
      Tracie JohnsonSep 2, 2025
      If linearity is out of the specified range, it could be caused by: * A defective tuning-fork sensor or mechanism unit – consider replacing it. * A mistake in the setting of address data – correct the setting. * Coefficient memories changed by noises or static electricity – adjust linearity. * Using an incorrect calibration weight – use the correct weight.
  • J
    jeremiah47Sep 5, 2025
    What to do if KERN EW - 600 - 2M Scales show too much corner error?
    • J
      Jeffrey VegaSep 5, 2025
      If there's too much corner error, it might be because: * The mechanism unit is defective, such as a bent or twisted Roberval plate (spring) – consider replacing the mechanism unit. * The pan base is touching other parts – adjust the position of the pan base.
  • J
    Jessica SmithSep 8, 2025
    What does 'b-Err' mean on KERN Scales?
    • R
      Ronald RodriguezSep 8, 2025
      A 'b-Err' message indicates that: * The EW/EG board is defective – consider replacing it. * Coefficient memories (address data) have changed due to noises or static electricity – adjust linearity.
  • D
    diana61Sep 12, 2025
    What does '1-Err' mean on KERN Scales?
    • K
      Kenneth SmithSep 12, 2025
      A '1-Err' message indicates that the calibration weight is less than ½ F.S. Use a calibration weight that is at least ½ F.S.

Summary

Total View of KERN EW;WG Balances

KERN EW;EG Electronic Construction

KERN EW Type Block Diagram

Illustrates the electronic components and connections for the KERN EW model.

KERN EG Type Block Diagram

Shows the block diagram for the electronic construction of the KERN EG model.

KERN EW Type Wiring Diagram

Details the complete wiring schematic for the KERN EW series balances.

KERN EG Type Wiring Diagram

Presents the overall wiring layout for the KERN EG precision balances.

KERN EW;EG Troubleshooting Procedures

Troubleshooting Procedure Guide

Step-by-step guide to diagnose and resolve issues with the KERN EW/EG balances.

Troubleshooting Symptom Table

Lists common symptoms and their potential causes and remedies for EW/EG balances.

Initial Diagnostic Checks

Essential primary checks to perform before detailed troubleshooting of the scale.

Removing the Scale Case

Instructions on how to safely remove the upper and lower cases of the balance.

Lock Switch Operation

Explanation of the ON/OFF lock switch function and its impact on settings.

Reassembling the Scale Case

Procedure for correctly covering the balance after servicing or adjustments.

KERN EW;EG Adjustments and Settings

Span Calibration (EW Type)

Procedure for performing span calibration on KERN EW series balances.

Corner Error Adjustment

Steps to adjust corner errors using adjustment bolts on the balance mechanism.

Overload Stopper Adjustment

Guide on how to set and adjust the overload stoppers for protection.

Built-In Weight Calibration (EG Type)

Procedure for calibrating the internal reference weight on EG series scales.

KERN EW;EG Parts Replacement

Mechanism Unit Replacement Sequence

Step-by-step guide for replacing the weighing mechanism unit.

Parts List for KERN EW Type

Parts List for KERN EG Type

KERN EW - 600 - 2M Specifications