EasyManua.ls Logo

KERN EW - 600 - 2M - KERN EW;EG Adjustments and Settings; Span Calibration (EW Type)

KERN EW - 600 - 2M
26 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EW/EG-KB-e-9811 13
4 SECTION 4. ADJUSTMENTS AND SETTINGS
4.1 Span calibration (Cal) - EW type
1. It is recommended to use a calibration weight of
better accuracy than the scale.
2. The calibration is available with ½ of F.S.
Nevertheless, it is recommended to use weight closer
to F.S. for accurate calibration.
3. Error messages:
0 - E r r : The calibration weight is over the full capacity.
1 - E r r : The calibration weight is less than ½ of the capacity.
2 - E r r : The data error exceeds 0.4%.
Or perhaps the scale may be defective.