10 MPE_HNM / MPE_PNM -SH-e-1810
5 The menu
Access to service menu „tCH“ is locked in verified weighing scales.
To disable the access lock, destroy the seal and actuate the
adjustment switch. For position of adjustment switch, see operating
instructions corresponding to the device chap. 16.
Attention:
After destruction of the seal the weighing system must be re-verified
by an authorised agency and a new verification wire/seal mark fitted
before it can be reused for applications subject to verification.
5.1 Navigation in the menu
In weighing mode, press and the first function [F1
oFF] will be displayed.
With help of , the individual functions can be
selected one after the other.
Confirm selected function by pressing .The current
setting will be displayed.
To select the desired setting, press and confirm
your selection by ; the weighing scale will return to
the menu.
Exit menu/
Return to weighing
mode
Press and the scales will return to weighing mode.