EasyManua.ls Logo

KERN MPE 250K100HNM - Adjustment Procedure

KERN MPE 250K100HNM
24 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MPE_HNM / MPE_PNM -SH-e-1810 17
7.1 Procedure:
In weighing mode, press repeatedly until [tCH] appears.
Press and [Pin] will appear.
Press , and one after the other and [P1 SPd] will
appear
Press , „P2 CAL“ will be displayed
Operate adjustment switch; for position see chap.16
Press and [dESC] will appear
Press repeatedly until „CAL“ will be displayed.
To confirm, press and [UnloAd] will appear
Ensure that there are no objects on the weighing pan.
Wait until the stability display „STABLE“ is displayed and then
confirm by pressing .
(example)
The size of the currently set adjustment weight will be
displayed.
To change, select the digit to be altered by , and the
numerical value by .
Confirm by pressing and [LoAd] will be displayed.