The displayed weight does not glow.
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
Power supply interrupted.
The displayed weight is permanently
changing
The weighing plate is in contact with
foreign matter.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
The weighing value is obviously
wrong
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Weighing plate has contact with other objects
Check ambient conditions (draught, vibrations etc.)
Internal adjustment automatics defective
Incorrect adjusting weight
Divergence last external adjustment > 1%
During the adjustment there was one weight on the weighing plate
Divergence from last internal adjustment > 1%
Insufficient battery capacity for adjustment
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.