PES/PEJ-BA-e-1316
90
Fault
Possible cause
The displayed weight does not glow.
The balance is not switched on.
The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
Power supply interrupted.
The displayed weight is permanently
changing
Draught/air movement
Table/floor vibrations
The weighing plate is in contact with
foreign matter.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
The weighing value is obviously
wrong
The display of the balance is not at zero
Adjustment is no longer correct.
Great fluctuations in temperature.
Electromagnetic fields / static charging
(choose different location/switch off inter-
fering device if possible)
Error message
Possible cause
o-Err
Weighing range exceeded
u-Err
Weighing plate has contact with other objects
b-Err
Check ambient conditions (draught, vibrations etc.)
d-Err
Damaged electronics
A-Err
Internal adjustment automatics defective
1-Err
Incorrect adjusting weight
2-Err
Divergence last external adjustment > 1%
3-Err
During the adjustment there was one weight on the weighing plate
4-Err
Divergence from last internal adjustment > 1%
7-Err
Insufficient battery capacity for adjustment
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.

Table of Contents

Other manuals for KERN PES 620-3M

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the KERN PES 620-3M and is the answer not in the manual?

KERN PES 620-3M Specifications

General IconGeneral
BrandKERN
ModelPES 620-3M
CategoryScales
LanguageEnglish

Related product manuals