34 PFB-BA-e-1630
7.5 Percent weighings
Percent weighing allows to display weight in percent, in relation to a reference
weight.
Percent weighing allows to display weight in percent, in relation to a reference
weight.
7.5.1 Non-verifiable models
Balance display
Operation
Start balance by pressing
The balance will carry out a self-test.
Wait for the „0.0 g“ display.
Put a reference weight on the weighing plate, which
corresponds to 100%
Press
In the display appears100.00%
Remove reference weight
The display returns to0.0 %
Put on specimen
In the display appears the percentage of the specimen
with reference to the reference weight.
Back to weighing mode by pressing again the
button.