TALJG_A/TALSG_A/TPLJG_A/TPLSG_A-BA-e-2112 37
8.3 Adjustment using the internal weight after the Model CAL
button is pressed (models ALJ/PLJ)
Preliminary condition: Menu setting “Internal adjustment”, see chapter 8.1.
In the weighing mode, press the CAL button, the adjustment will be carried out
automatically.
,
After the successful adjustment, the scale will switch to the weighing mode
again automatically.
If an adjustment error occurs (e.g. following an impact), the error message
“CAL bUt” will appear on the display. Start the adjustment process again, by
pressing the CAL button.