Cautions:
Make sure that the code number shown on the screen is the same as the code number
printed on the Cholesterol test strip vial, otherwise you may get incorrect measurement.
1.
2.
3.
Steps for Running a Cholesterol Control Check
CAUTIONS:
Always check the Cholesterol code key number to make sure it matches the number
labeled on the test strip vial when you run a Cholesterol control test. Otherwise you may
get incorrect results.
Always write down the opening date of your Cholesterol control solution vial and test strip
vial. Both of them are good for only 2 months after first opening or until the expiration date,
whichever comes first.
The result will be stored in memory and appear just like the blood cholesterol result.
Insert the cholesterol code key into the code key slot on the back of the meter.
Take one strip from the vial. Close the vial quickly.
Insert the test strip into the test strip slot on the meter.
Themeter will first display code number ,
then the bloodsymbol along with date
.
CHOL
Code
9005
CHOL