Cautions:
Make sure that the code number shown on the screen is the same as the code number
printed on the Hemoglobin test strip vial, otherwise you may get incorrect measurement.
1.
2.
3.
Steps for Running a Hemoglobin Control Check
CAUTIONS:
Always check the Hemoglobin code key number to make sure it matches the number
labeled on the test strip vial when you run a Hemoglobin control test. Otherwise you may
get incorrect results.
Always write down the opening date of your
Hemoglobin control
solution vial and test strip
vial. Both of them are good for only 2 months after first opening or until the expiration date,
whichever comes first.
The result will be stored in memory and appear just like the blood cholesterol result.
Insert the hemoglobin code key into the code key slot on the back of the meter.
Take one strip from the vial. Close the vial quickly.
Insert the test strip into the test strip slot on the meter.
The meter will rst display code number ,
then the blood symbol along with date
.
Hb
Code
8702
Hb
pm
17

Table of Contents