4.
Hold the Hemoglobin control solution bottle upside down,slowly
squeeze the bottle to form a small drop, discard the first drop of the
solution, let the second drop touch the edge of the sample target area
on the test strip.The control solution will be drawn into the reaction zone
automatically.
5.
You will hear a “Beep” sound. Close the cap of the control solution
immediately.The meter starts to count down 10 seconds then show
your result on the screen.
6.
Check if the displayed
Hemoglobin
value falls within
the
acceptable range shown on the test strip vial.
7.
Remove the test strip from the meter and discard it.