2Quick Start
44 Keysight E3633A and E3634A Service Guide
Adjust OVP trip level
1Raise the OVP trip level.
Press key and turn the knob to raise the OVP trip level.
2 Move to the OVP CLEAR mode.
Press key to move the OVP clear mode. The “OVP ON” message
appears on the display. Turn the knob to the right until the above message
appears on the display.
3 Clear the overvoltage cond ition and exit this menu.
Now, when you press key again, the “DONE’’ message is displayed for
a second and the OVP annunciator will not blink any more. The output will
return to the meter mode.
OVP CLEAR