Calibration Procedures 3
Keysight E3633A and E3634A Service Guide 71
Performance Test Record for Keysight E3633A and E3634A
CV performance test record
Test Description Models
Actual
Result
Specifications
Upper Limit Lower Limit
CV Programming Accuracy @ 0 volts
(DVM reading)
Both +0.0100 V -0.0100 V
CV Readback Accuracy @ 0 volts Both DVM + 0.0050 V DVM - 0.0050 V
CV Programming Accuracy
@ Full Scale (DVM reading)
(E3633A) +20.0200 V 19.9800 V
(E3634A) +50.0350 V 49.9650 V
CV Readback Accuracy @ Full Scale
(E3633A) DVM + 0.0150 V DVM - 0.0150 V
(E3634A) DVM + 0.0300 V DVM - 0.0300 V
CV Load Effect (Load Regulation)
Maximum change: < (4 mV)
[a]
or (7 mV)
[b]
[a] For Keysight E3633A Model
[b] For Keysight E3634A Model
CV Source Effect (Line Regulation)
Maximum change: < (4 mV)
[a]
or (7 mV)
[b]
CV PARD (Normal mode)
< 3 mV p-p, (350
μ
V)
[a]
or (500
μ
V)
[b]
rms
Load Transient Response Time Both
< 50
μ
sec