2Quick Start
48 Keysight E3633A and E3634A Service Guide
Adjust OCP trip level
1 Raise the OCP trip level.
Press key and turn the knob to raise the OCP trip level.
2 Move to the OCP CLEAR mode.
Press the key to move to the OCP CLEAR mode. The “OCP ON” message
appears on the display. Turn the knob to the right until the above message
appears on the display.
3 Clear the overcurrent condition and exit this menu.
Now, when you press again, the “DONE” message is displayed for a
second and the OCP annunciator will not blink any more. The output will
return the meter mode.
OCP CLEAR