3 Calibration Procedures
60 Keysight E3633A and E3634A Service Guide
4 Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limit of (0 V ± 10 mV). Also, note that the CV, Adrs, Limit,
and Rmt annunciators are on.
5 Readback the output voltage over the remote interface by sending the
command:
6 Record the value displayed on the controller. This value should be within the
limit of (DVM ±5 mV).
7 Program the output voltage to full rated value (20.0 V)
[1]
or (50.0 V)
[2]
by
sending the command:
8 Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limit of (20 V ± 20 mV)
[1]
or (50 V ± 35 mV)
[2]
9 Readback the output voltage over the remote interface by sending the
command:
10 Record the value displayed on the controller. This value should be within the
limit of (DVM ± 15 mV)
[1]
or (DVM ± 30 mV)
[2]
.
MEAS:VOLT?
[1] For Keysight E3633A Model
[2] For Keysight E3634A Model
VOLT 20.0 (E3633A)
VOLT 50.0 (E3634A)
MEAS:VOLT?

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies E3633A and is the answer not in the manual?

Keysight Technologies E3633A Specifications

General IconGeneral
BrandKeysight Technologies
ModelE3633A
CategoryPower Supply
LanguageEnglish

Related product manuals