Remote Interface Reference 3
E364xA User’s and Service Guide 123
Table 3-2 Power supply state
[a]
[a] The voltage and current step sizes listed are typical values.
Command E3646A
state
E3647A
state
E3648A
state
E3649A
state
CURR 3 A0.8 A5 A1.4 A
CURR:STEP 0.052 mA 0.015 mA 0.095 mA 0.026 mA
CURR:TRIG 3 A 0.8 A 5 A 1.4 A
DISP ON
OUTP OFF
OUTP:REL OFF
TRIG:DEL 0
TRIG:SOUR BUS
VOLT 0 V
VOLT:STEP 0.35 mV 1.14 mV 0.38 mV 1.14 mV
VOLT:TRIG 0 V
VOLT:PROT 22.0 V 66.0 V 22.0 V 66.0 V
VOLT:PROT:STAT ON
VOLT: RANG P8V (Low) P25V (Low) P8V (Low) P35V (Low)

Table of Contents