1 Getting Started
50 E364xA User’s and Service Guide
Multiple loads
When connecting multiple loads to the power supply, each load should be
connected to the output terminals using separate connecting wires. This
minimizes mutual coupling effects between loads and takes full advantage of the
low output impedance of the power supply. Each pair of wires should be as short
as possible and twisted or bundled to reduce lead inductance and noise pick- up.
If a shield is used, connect one end to the power supply ground terminal and leave
the other end disconnected.
If cabling considerations require the use of distribution terminals that are located
remotely from the power supply, connect output terminals to the distribution
terminals by a pair of twisted or shielded wires. Connect each load to the
distribution terminals separately.