EasyManua.ls Logo

Keysight Technologies E3649A

Keysight Technologies E3649A
255 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service and Maintenance 8
E364xA User’s and Service Guide 233
General Calibration or Adjustment Procedure
The front panel calibration procedures are described in this section.
For voltage calibration, disconnect all loads from the power supply and
connect a digital voltmeter (DVM) across the output terminals.
For current calibration, also disconnect all loads from the power supply,
connect an appropriate current monitoring resistor 0.01
across the output
terminals, and connect a DVM across the terminals of the monitoring resistor.
You can abort a calibration at any time by turning the power supply off from
the front panel, by issuing a remote interface device clear message, or by
pressing the front panel Local key.
Table 8-8 shows the calibration parameters and points which should be used to
calibrate the output voltage and current.
NOTE
The power supply should be calibrated after a 1-hour warm-up with no load
connected. Perform the voltage calibration prior to the OVP calibration.
Table 8-8 Parameters for calibration
Output Calibration parameter Voltage/ Current Calibration point
mnemonic
Output 1 VOLT1 CAL Voltage V LO
V MI
V HI
OVP1 CAL OVP None
CURR1 CAL Current I LO
I MI
I HI

Table of Contents

Related product manuals