Characteristic and Specifications 7
E364xA User’s and Service Guide 195
Overvoltage protection
accuracy
±(% of output + offset)
<0.5% + 0.5 V
Overvoltage protection
response time
<1.5 ms when the trip voltage is .3 V
<10 ms when the trip voltage is <3 V
Input power option
(selectable via two switches
on the PC board)
115 Vac ±10%, 47 Hz to 63 Hz (standard)
230 Vac ±10%, 47 Hz to 63 Hz (option 0E3)
100 Vac ±10%, 47 Hz to 63 Hz (option 0E9)
Maximum input power 105 VA 105 VA 205 VA 175 VA
[a] Maximum change in the output or readback per °C after a 30-minute warm-up.
[b] Following a 30-minute warm-up, with the output in the ON state according to the operating mode (CC with load or CV), and with a change in
the output over 8 hours under constant load, line, and ambient temperature.
[c] During turn-on or turn-off of the AC power, if the output control is set to less than 1 V.
Table 7-3 Supplemental characteristics