EasyManua.ls Logo

Keysight Technologies E3649A

Keysight Technologies E3649A
255 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Service and Maintenance 8
E364xA User’s and Service Guide 227
1 Turn off the power supply and connect the output to be tested as shown in
Figure 8-6 with the current monitoring resistor 0.2
(R
M2
) across output
terminals. Connect an rms voltmeter across the current monitoring resistor as
shown in Figure 8-6.
2 Turn on the power supply. Select the output to be tested and the low voltage
range (8 V/3 A)
[1]
, enable the output, and set the display to the limit mode.
When the display is in the limit mode, program the current to the full-scale
rated (3 A)
[1]
and the voltage to the maximum programmable voltage (8.24
V)
[2]
.
3 Divide the reading on the rms voltmeter by the load resistance to obtain the
rms current. The readings should be within the limit of 4 mA.
4 Repeat step 1 through step 3 for the other output.
[1] For the E3646A model. For other models, refer to Table 8-5.
[2] For the E3646A model. For other models, refer to Table 7-3.

Table of Contents

Related product manuals