Step 2 Run the input sequence
In this example, the load operation is in CC mode.
Set I
1
, trigger delay and repeat count accordingly: Use the navigation keys to select the field; use the numeric
entry keys to enter the value. The value is set when you press [Enter].
Set Trigger Source to Key. Press Trig Src Key.
Press the color-coded [On] key to enable the selected input.
Press Run to start the sequence operation. To abort the operation, press Stop.
Press Back to exit and return to the previous menu.
From the remote interface:
To configure the load operating mode to CC mode:
FUNC CURR, (@1)
To set the sequencer to Toggle mode:
TRAN:MODE TOGG, (@1)
To configure the main current level with 5 A:
CURR 5, (@1)
To configure the value for the transient current level with 10 A:
CURR:TLEV 10, (@1)
To set the trigger source of input 1 to Key/Immediate:
TRIG:TRAN:SOUR IMM, (@1)
To configure the input voltage to follow transient value when a trigger occurs:
CURR:MODE LIST
To enable input 1:
INP ON, (@1)
To initiate the transient operation sequence:
INIT:TRAN, (@1)