EasyManua.ls Logo

Keysight Technologies EL30000 Series - Step 2 - Run the Input Sequence

Keysight Technologies EL30000 Series
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameter Available Key
Settings
Description
Trigger Source Key, IO, or Rmt Sets the trigger source for the system.
Default is Key.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (List Run/Stop Key) Selects the Run Stopped softkey as a trigger
source.
IO (DIO Trigger In) Selects any configured digital IO (with Trigger
Input function) as a trigger source
Rmt (Remote Command) Selects a remote interface command as a trigger
source.
Trigger Delay 0 to 0.255 s Sets the trigger delay in seconds.
Default is 0 s.
Step 2 Run the input sequence
In this example, the load operation is in CV mode.
Set V
1
, pulse width, trigger delay and repeat count accordingly: Use the navigation keys to select the field; use the
numeric entry keys to enter the value. The value is set when you press [Enter].
Set Trigger Source to Key. Press Trig Src Key.
Press the color-coded [On] key to enable the selected input.
Press Run to start the sequence operation. To abort the operation, press Stop.
Press Back to exit and return to the previous menu.
Keysight EL30000 Series User's Guide 97

Table of Contents

Related product manuals