Parameter Available Key Settings Description
Trigger Source Key, IO, or Rmt Sets the trigger source for the system.
Default is Key.
Action required:
Press Trig Src to toggle between Key, IO, and Rmt.
Key (List Run/Stop Key) Selects the Run Stopped softkey as a trigger
source.
IO (DIO Trigger In) Selects any configured digital IO (with Trigger
Input function) as a trigger source
Rmt (Remote Command) Selects a remote interface command as a trigger
source.
Trigger Delay 0 to 0.255 s Sets the trigger delay in seconds.
Default is 0 s.
Repeat Count 1 to 9999 Sets the list repeat count. This sets the number of times that a list is executed before it
completes.
Default is 1.
Continuous - Enables check box to repeat the list continuously.
Step 2 Run the input sequence
In this example, the load operation is in CV mode.
Set V
1
, period, frequency, duty cycle, trigger delay and repeat count accordingly: Use the navigation keys to
select the field; use the numeric entry keys to enter the value. The value is set when you press [Enter].
Set Trigger Source to Key. Press Trig Src Key.
Press the color-coded [On] key to enable the selected input.
Press Run to start the sequence operation. To abort the operation, press Stop.
Press Back to exit and return to the previous menu.
94
Keysight EL30000 Series User's Guide