Utilities Menu - Test / Setup
Test / Setup provides access to calibration, self-test, as well as configures user preferences, sets the date and time,
and accesses the help:
MEMORYSANITIZATION PROCEDURE
Refer to the
EL30000 Series Security Guide
for instructions to perform a memory sanitization procedure that san-
itizes all user-accessible instrument memory and restarts the instrument.
Calibration
Calibration accesses the instrument calibration procedure.
The default security passcode is 0. Press Login to access the calibration options.
Perform Calibration opens the calibration menu. Make sure you have run the self-test before performing any
calibration. Refer to "Calibration Adjustment Procedures" in the Service Guide for details.
Press Change Passcode to change the passcode using the numeric keypad. The passcode can be set up to 9 digits.
Auto save On will automatically save the calibration data upon logout. When Auto save is Off, press Cal Save to
save the calibration data.
Self-Test
Self Test verifies proper instrument operation.
Keysight EL30000 Series User's Guide 133