Environmental Conditions
The EL30000 Series is designed for indoor use. The table below shows the general environmental requirements for
this instrument.
Environmental condition Requirement
Temperature Operating condition: 0 °C to 40 °C
Storage condition: 40 °C to 70 °C
Humidity Operating condition: Up to 85% RH at temperature up to 40 °C (non-condensing)
Storage condition: Up to 90% RH (non-condensing)
Altitude Up to 2000 m
Pollution degree 2
Installation category II (for AC input)
Mains voltage (rms) 100 VAC to 240 VAC ±10%
Mains frequency 50/60 Hz
Acoustic noise declaration Sound pressure: < 65 dB(A) at operator position, Lρ < 70 dB(A) at bystander position
Sound power: Lw < 70 dB(A)