Keysight FieldFox Handheld Analyzers Service Guide 125
Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
4. Carefully remove the serial number label from the inside bottom of the old
front case and apply the label to the new front case.
5. Refer to Figure 7-13. Choose the appropriate model number nameplate
overlay for your analyzer, remove the protective backing to expose the
adhesive, and adhere it in the location shown.
6. Remove the dowel pins (5) from the old front case and install them in the
new case. The other two dowel pins should already be installed in the
case.
7. Make sure that the gaskets (6) are all in place in the grooves, that there
are no gaps, and that they are not damaged in any way.
8. Refer to “Removing and Replacing the Front Panel Group Assemblies” on
page 130 to reinstall the front panel group assemblies removed in step 3.
9. Refer to “Removing and Replacing the Measurement Group Assemblies”
on page 136 to reinstall the measurement group assemblies removed in
step 2.
10.Reinstall the rear case onto the front case. Refer to “Combining the Front
and Rear Cases” on page 126.