Keysight FieldFox Handheld Analyzers Service Guide 41
Troubleshooting
FieldFox High-Temperature Protection
FieldFox High-Temperature Protection
The following features prevent degradation or damage in the event of high
internal temperatures in the FieldFox.
How to monitor the internal FieldFox temperature:
—Press System, then Service Diagnostics.
—Press Internal Temperatures.
The temperature at which the following events occur is the average of the
RF1, RF2, SB1, SB2 temperatures. These temperatures come from internal
sensors embedded within FieldFox.
Temperature Control Mode
At approximately 75°C, the FieldFox enters Temperature Control mode by
reducing display intensity, switching to Outdoor Sun display colors, and
reducing measurement speed. This should decrease the internal temperature
which preserves measurement accuracy and maintains the long-term reliability
of the FieldFox. When this occurs, the following message is displayed on the
FieldFox screen:
The system is entering Temperature Control Mode due to high
internal temperature.
When entering Temperature Control mode, save your instrument state and
data that you want to keep.
When the temperature drops to approximately 73°C, a message is displayed
indicating that the FieldFox is leaving Temperature Control Mode and normal
operating settings are restored.
High-Temp Shut down
In extreme situations, Temperature Control mode may not stop an increase in
the FieldFox internal temperature. At approximately 78°C, High-Temperature
Shut down will engage and turn OFF the FieldFox.
Just prior to shut down, the FieldFox will display a warning of imminent shut
down.
Do NOT store the FieldFox in the softcase while powered ON or in Standby mode.
Measurement speed specifications do NOT apply in Temperature Control Mode.

Table of Contents

Other manuals for Keysight Technologies FieldFox N9912A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies FieldFox N9912A and is the answer not in the manual?

Keysight Technologies FieldFox N9912A Specifications

General IconGeneral
BrandKeysight Technologies
ModelFieldFox N9912A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals