Keysight FieldFox Handheld Analyzers Service Guide 129
Repair and Replacement Procedures
Removing and Replacing the Case Group Parts
Figure 7-17 Combining the Front and Rear Cases (N995xA, N996xA)
13.For N995xA and N996xA models only. Refer to Figure 7-4. Place the nut
rings onto the RF connectors. Using a 1.5 mm Allen torque screwdriver,
tighten the 3 Allen screws on each nut ring. Torque to 24 ozf-in.
14.Reinstall the battery. Refer to “Removing and Replacing the Main Battery”
on page 109.