Reference 22
Keysight InfiniiVision 3000 X-Series Oscilloscopes User's Guide 327
Environmental Conditions
CAUTION
When measuring voltages over 30 V, use a 10:1 probe.
CAUTION
Maximum input voltage at digital channels
±40 V peak
Environment Indoor use only.
Ambient
temperature
Operating 5 °C to +55 °C; non-operating –40 °C to +71 °C
Humidity Operating: Up to 95% RH, non-condensing to temperatures up to +40 °C
decreasing linearly to 50% RH at +55 °C
Non-operating: Up to 95% RH up to +40 °C. Up to 45% RH up to +50 °C.
(non-condensing)
Altitude Operating: Up to 4,000 m
Non-operating: Up to 15,300 m
Overvoltage
Category
This product is intended to be powered by MAINS that comply to Overvoltage
Category II, which is typical of cord-and-plug connected equipment.
Pollution Degree The InfiniiVision 2000/3000 X-Series oscilloscopes may be operated in
environments of Pollution Degree 2 (or Pollution Degree 1).
Pollution Degree
Definitions
Pollution Degree 1: No pollution or only dry, non-conductive pollution occurs.
The pollution has no influence. Example: A clean room or climate controlled
office environment.
Pollution Degree 2. Normally only dry non-conductive pollution occurs.
Occasionally a temporary conductivity caused by condensation may occur.
Example: General indoor environment.
Pollution Degree 3: Conductive pollution occurs, or dry, non-conductive
pollution occurs which becomes conductive due to condensation which is
expected. Example: Sheltered outdoor environment.