568 N9030B PXA Signal Analyzer Service Guide
Post-Repair Procedures
Post-Repair Procedures
1. Press Enter to proceed with the Windows XP setup.
2. When you see the Windows XP Setup Wizard screen as seen in Figure
17-10, press Enter to move to the License Agreement screen.
Figure 17-10 Windows XP Setup Wizard
3. When the Windows XP License Agreement window comes up, select “I
accept this agreement” and press Next.
4. The Date and Time Settings screen is used to set the appropriate date,
time and time zone. These settings will be configured later by the end
user.
Select Next to continue the Windows XP Setup Wizard.
5. Wait for the Windows XP Setup Wizard to complete. The instrument will
restart itself a few times.
6. Once the Windows XP Setup Wizard has completed the instrument is
ready for the next step of configuration.