SYSTem:COMMunicate:LAN:NETBios {OFF|0|ON|1}
SYSTem:COMMunicate:LAN:NETBios?
Disables or enables the use of NETBios to automatically assign an IP
address to the 33210A for peer-to-peer naming on a private network that
uses the NETBios naming service. The factory setting is enabled (ON).
The query returns “0” (OFF) or “1” (ON).
If you change the NETBios setting, you must cycle power on the
instrument to activate the new setting.
The NETBios setting is stored in non-volatile memory, and does not
change when power has been off, or after an instrument reset (*RST).
SYSTem:COMMunicate:LAN:TELNet:PROMpt <
string
>
SYSTem:COMMunicate:LAN:TELNet:PROMpt?
Sets the command prompt seen when using a Telnet session to
communicate with the 33210A. The default prompt is 33210A>”.
The 33210A uses LAN port 5024 for Telnet sessions.
Telnet session can typically be started as follows from a host
computer shell:
telnet <IP_address> <port>
For example:
telnet 169.254.2.20 5024
To exit a Telnet session, press <Cntr-D>.
The prompt is a quoted string of up to 15 characters.
The command prompt is stored in non-volatile memory, and does not
change when power has been off, or after an instrument reset (*RST).