1 Performance Tests and Calibration
Calibration Overview
2 34450A Service Guide
Calibration Overview
.
Closed-case calibration
The instrument features closed- case electronic calibration. No internal
mechanical adjustments are required. The instrument calculates
correction factors based upon the input reference value you set. The new
correction factors are stored in non- volatile memory until the next
calibration adjustment is performed. Non- volatile EEPROM calibration
memory does not change when power has been off or after a remote
interface reset.
Keysight Technologies calibration services
When your instrument is due for calibration, contact your local Keysight
Service Center for a low- cost re- calibration. The 34450A is supported on
automated calibration systems, which allow Keysight to provide this
service at competitive prices.
NOTE
Make sure you have read “Test Considerations” on page 5 before calibrating the
instrument.