1 Performance Tests and Calibration
Recommended Test Equipment
4 34450A Service Guide
Recommended Test Equipment
The test equipment recommended for the performance verification and
adjustment procedures is listed in Table 1- 1 below. If the exact
instrument is not available, substitute calibration standards of equivalent
accuracy.
A suggested alternate method would be to use the Keysight 3458A 8½
Digit Digital Multimeter to measure less accurate yet stable sources. The
output value measured from the source can be entered into the
instrument as the target calibration value.
Table 1 - 1 Recommended test equipment
Application Recommended Equipment Recommended Accuracy Requirements
Zero calibration Fluke 5520A or shorting Plug-Dual banana plug with
copper wire short between the two terminals
DC voltage Fluke 5520A <1/5 instrument 1 year spec
DC current Fluke 5520A <1/5 instrument 1 year spec
Resistance Fluke 5520A <1/5 instrument 1 year spec
AC voltage Fluke 5520A <1/5 instrument 1 year spec
AC current Fluke 5520A <1/5 instrument 1 year spec
Frequency Fluke 5520A <1/5 instrument 1 year spec
Capacitance Fluke 5520A <1/5 instrument 1 year spec