1 Performance Tests and Calibration
Calibration Process
26 34450A Service Guide
Calibration Process
The following general procedure is the recommended method to complete
a full instrument calibration.
1 Read “Test Considerations” on page 5.
2 Perform the verification tests to characterize the instrument (incoming
data).
3 Unsecure the instrument for calibration (see “Calibration Security” on
page 23).
4 Perform the adjustment procedures (see “Adjustments” on page 29).
5 Secure the instrument against calibration.
6 Note the new security code and calibration count in the instrument's
maintenance records.
NOTE
The new security code and calibration count in the instrument’s maintenance records.