1 Performance Tests and Calibration
Performance Verification Tests
18 34450A Service Guide
1 V 10 kHz 1 V ±3 mV
1 V 30 kHz 1 V ±18 mV
1 V 100 kHz 1 V ±33 mV
10 V 20 Hz 10 V ±0.11 V
10 V 45 Hz 10 V ±30 mV
10 V 1 kHz 10 V ±30 mV
10 V 10 kHz 10 V ±30 mV
10 V 30 kHz 10 V Q ±0.18 V
10 V 100 kHz 10 V ±0.33 V
100 V 45 Hz 100 V ±0.3 V
100 V 1 kHz 100 V ±0.3 V
100 V 10 kHz 100 V Q ±0.3 V
100 V 30 kHz 100 V ±1.8 V
100 V 100 kHz 100 V ±3.3 V
750 V 45 Hz 750 V ±2.25 V
750 V 1 kHz 750 V ±2.25 V
750 V 10 kHz 750 V ±2.25 V
[1] Select Slow Mode 5½ digit measurement resolution
Q= Quick performance verification test points
Table 1 - 7 AC volts verification test
Vrms Input Frequency Function
[1]
Range Quick Check
Error from
Nominal 1 Year