CALCulate:SMOothing[:STATe] {OFF|ON}
CALCulate:SMOothing[:STATe]?
Enables or disables a smoothing (moving average) filter on the measurements. Smoothing is intended to
average out small variations in the measurements. Larger variations will cause the filter to reset.
Parameter Typical Return
{ON|1|OFF|0}. Default: ON. 0(OFF)or1(ON)
Enable a 50-point moving average filter on 1000 measurements:
CONF:VOLT 10
SAMP:COUN 1000
CALC:SMO:RESP MED
CALC:SMO:STAT ON
CALC:STAT ON
READ?
Typical Response:+9.99383828200000E+03,+9.99352315400000E+03,...
(1000 measurements)
l Applies only to the 34465A and 34470A.
l The CALCulate subsystem and the individual calculate functions are disabled for timestamp meas-
urements.
l The instrument disables the smoothing (moving average) filter after a Factory Reset (*RST) or an Instru-
ment Preset (SYSTem:PRESet).
Keysight Truevolt Series Operating and Service Guide
235
SCPI Programming Reference